Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours or 4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (200 uL), Serum (200 uL), EDTA Plasma (200 uL), Heparin Plasma (200 uL), Citrate Plasma (200 uL)
Sensitivity
3.9 pg/mL
Assay Range
7.8 - 500 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human IL-10
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 24.4 | 79.6 | 223 | 23.6 | 74.2 | 227 |
| Standard Deviation | 1.6 | 3.9 | 5.5 | 1.8 | 4.9 | 12.7 |
| CV% | 6.6 | 4.9 | 2.5 | 7.6 | 6.6 | 5.6 |
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 23.9 | 76.9 | 231 | 23.2 | 75.1 | 228 |
| Standard Deviation | 1.2 | 3.3 | 3.9 | 1.7 | 5.6 | 13.4 |
| CV% | 5 | 4.3 | 1.7 | 7.3 | 7.5 | 5.9 |
Recovery
The recovery of human IL-10 spiked to three levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 98 | 88-110 |
| Citrate Plasma (n=5) | 96 | 87-108 |
| EDTA Plasma (n=5) | 92 | 75-105 |
| Heparin Plasma (n=5) | 94 | 84-105 |
| Serum (n=5) | 100 | 92-110 |
Linearity
To assess the linearity of the assay, five samples were spiked with high concentrations of IL-10 in various matrices and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.