Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
2.5 hours
Sample Type & Volume Required Per Well
Serum (20 uL), EDTA Plasma (20 uL), Heparin Plasma (20 uL), Citrate Plasma (20 uL)
Sensitivity
0.5 nmol/L
Assay Range
3.0 - 80 nmol/L (Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Measures human sTfR levels in various samples
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
Interference observed with other substances.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 60 | 60 | 60 |
| Mean | 6.9 | 18.2 | 40.2 | 10.9 | 25.7 | 67.2 |
| Standard Deviation | 0.3 | 1.3 | 2.5 | 0.7 | 1.4 | 3.8 |
| CV% | 4.3 | 7.1 | 6.2 | 6.4 | 5.4 | 5.7 |
Recovery
The recovery of sTfR spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Citrate Plasma (n=5) | 112 | 105-121 |
| EDTA Plasma (n=5) | 103 | 97-112 |
| Heparin Plasma (n=5) | 106 | 101-112 |
| Serum (n=5) | 97 | 95-100 |
Linearity
To assess the linearity of the assay, five samples containing high concentrations of sTfR were diluted with Specimen Diluent and evaluated.