返回 产品中心
Human IL-8/CXCL8 Quantikine ELISA Kit
Human IL-8/CXCL8 Quantikine ELISA Kit Summary
Catalog Size / Price
D8000C 1 Kit / Please InquireS8000C 1 Pack (6 Plates) / Please Inquire
PD8000C 1 Pack (50 Plates)
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL), Citrate Plasma (50 uL)
Sensitivity
7.5 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human IL-8
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) hree samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 168 | 526 | 1093 | 196 | 581 | 1160 |
| Standard Deviation | 9.4 | 28.6 | 70.7 | 14.5 | 56.6 | 70.2 |
| CV% | 5.6 | 5.4 | 6.5 | 7.4 | 9.7 | 6.1 |
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 115 | 386 | 802 | 132 | 410 | 817 |
| Standard Deviation | 5.3 | 17 | 37.9 | 10.7 | 28 | 42.4 |
| CV% | 4.6 | 4.4 | 4.7 | 8.1 | 6.8 | 5.2 |
Recovery
The recovery of IL-8 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 98 | 85-114 |
| Citrate Plasma (n=5) | 105 | 95-114 |
| EDTA Plasma (n=5) | 103 | 97-111 |
| Heparin Plasma (n=5) | 102 | 92-107 |
| Serum (n=5) | 98 | 88-106 |
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of IL-8 in various matrices and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有
Copyright © 深圳子科生物科技有限公司版权所有