返回 产品中心

Human CD14 Quantikine ELISA Kit

    

Human CD14 Quantikine ELISA Kit Summary

Catalog #  Size 

DC140     1 Kit 

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL), Citrate Plasma (10 uL)
Sensitivity
125 pg/mL
Assay Range
250.0 - 16,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human sCD14
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

Product Summary

The Quantikine Human soluble CD14 Immunoassay is a 4.5 hour solid phase ELISA designed to measure human soluble CD14 in cell culture supernates, serum, and plasma. It contains recombinant human CD14 expressed from CHO cells and antibodies raised against the recombinant factor. This immunoassay has been shown to accurately quantitate the recombinant factor. Results obtained measuring natural human sCD14 showed dose-response curves that were parallel to the standard curves obtained using the kit standards. These results indicate that this kit can be used to determine relative mass values for natural human sCD14.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 40 40 40
Mean 1111 2158 4187 1159 2298 4346
Standard Deviation 71 104 216 86 153 209
CV% 6.4 4.8 5.2 7.4 6.7 4.8

Recovery

The recovery of sCD14 spiked to three different levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=5) 91 85-105
Citrate Plasma (n=5) 102 88-110
EDTA Plasma (n=5) 98 90-107
Heparin Plasma (n=5) 96 91-104
Serum (n=5) 98 84-107

Linearity

To assess the linearity of the assay, five samples containing or spiked with high concentrations of sCD14 were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Human CD14 ELISA Linearity