返回 产品中心
Human Free BDNF Quantikine ELISA Kit
Human Free BDNF Quantikine ELISA Kit Summary
货号:DBD00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), EDTA Plasma (50 uL), Platelet-poor Heparin Plasma (50 uL), Citrate Plasma (50 uL)
Sensitivity
20 pg/mL
Assay Range
62.5 - 4,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Platelet-poor Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human BDNF
Cross-reactivity
Cross-reactivity observed with 1 or more available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
Interference observed with 1 or more available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Platelet-poor Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 476 | 1258 | 2175 | 528 | 1338 | 2287 |
| Standard Deviation | 24 | 48.2 | 135.7 | 59.7 | 102.3 | 186 |
| CV% | 5 | 3.8 | 6.2 | 11.3 | 7.6 | 8.1 |
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 334 | 825 | 1339 | 358 | 905 | 1517 |
| Standard Deviation | 15.6 | 28.3 | 56.5 | 36.8 | 86.8 | 121.6 |
| CV% | 4.7 | 3.4 | 4.2 | 10.3 | 9.6 | 8.1 |
Recovery
The recovery of BDNF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 96 | 83-105 |
| Platelet-poor Citrate Plasma (n=5) | 99 | 87-120 |
| Platelet-poor EDTA Plasma (n=5) | 96 | 89-115 |
| Platelet-poor Heparin Plasma (n=5) | 93 | 81-119 |
| Serum (n=5) | 100 | 85-116 |
Linearity
To assess the linearity of the assay, five samples containing or spiked with high concentrations of BDNF in various matrices were diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.