返回 产品中心
Human sVCAM-1/CD106 Quantikine ELISA Kit
Human sVCAM-1/CD106 Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DVC00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
2 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (10 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL)
Sensitivity
1.26 ng/mL
Assay Range
6.3 - 200 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human sVCAM-1
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 583 | 1423 | 2421 | 670 | 1571 | 2726 |
| Standard Deviation | 13.6 | 49.5 | 86.4 | 52.1 | 120.9 | 151.2 |
| CV% | 2.3 | 3.5 | 3.6 | 7.8 | 7.7 | 5.5 |
Linearity
To assess the linearity of the assay, samples spiked with high concentrations of sVCAM-1 were serially diluted with the Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705