Endothelin-1 Quantikine ELISA Kit
Endothelin-1 Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DET100
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (75 uL), Serum (75 uL), EDTA Plasma (75 uL), Heparin Plasma (75 uL), Urine (75 uL)
Sensitivity
0.207 pg/mL
Assay Range
0.4 - 25 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine)
Specificity
Synthetic and natural Endothelin-1
Cross-reactivity
Cross-reactivity observed with 1 or more available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 3 | 7.34 | 14.7 | 3.05 | 7.43 | 14.4 |
| Standard Deviation | 0.12 | 0.17 | 0.28 | 0.231 | 0.438 | 0.759 |
| CV% | 4 | 2.3 | 1.9 | 7.6 | 5.9 | 5.3 |
Recovery
The recovery of Endothelin-1 spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 99 | 88-108 |
| Human EDTA Plasma (n=4) | 93 | 87-102 |
| Human Heparin Plasma (n=4) | 93 | 85-106 |
| Human Serum (n=4) | 98 | 86-107 |
| Mouse EDTA Plasma (n=4) | 99 | 102-107 |
| Mouse Serum (n=4) | 94 | 86-104 |
| Rat EDTA Plasma (n=4) | 99 | 86-113 |
| Rat Serum (n=4) | 98 | 88-113 |
| Urine (n=4) | 91 | 85-109 |
Linearity
To assess the linearity of the assay, samples spiked with high concentrations of Endothelin-1 were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.