返回 产品中心
Human Galectin-3 Quantikine ELISA Kit Summary
Human Galectin-3 Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DGAL30
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL)
Sensitivity
0.085 ng/mL
Assay Range
0.3 - 10 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human Galectin-3. This assay cross-reacts 58% with Mouse Galectin-3.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity observed with 1 or more species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 0.79 | 2.46 | 5.11 | 0.8 | 2.59 | 5.5 |
| Standard Deviation | 0.03 | 0.09 | 0.18 | 0.05 | 0.16 | 0.32 |
| CV% | 3.8 | 3.7 | 3.5 | 6.3 | 6.2 | 5.8 |
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 0.78 | 2.27 | 5.06 | 0.81 | 2.51 | 5.29 |
| Standard Deviation | 0.03 | 0.1 | 0.15 | 0.07 | 0.2 | 0.36 |
| CV% | 3.8 | 4.4 | 3 | 8.6 | 8 | 6.8 |
Recovery
The recovery of Galectin-3 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 102 | 96-106 |
| EDTA Plasma (n=4) | 105 | 93-113 |
| Heparin Plasma (n=4) | 103 | 95-110 |
| Serum (n=4) | 104 | 94-115 |
Linearity
To assess the linearity of the assay, samples containing or spiked with high concentrations of human Galectin-3 in various matrices were diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705