返回 产品中心
Human CXCL1/GRO alpha Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human CXCL1/GRO alpha Quantikine ELISA Kit Summary
货号:DGR00B 规格:96T
Assay Length
4.25-4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (200 uL), Serum (200 uL), Platelet-poor EDTA Plasma (200 uL), Platelet-poor Heparin Plasma (200 uL), Platelet-poor Citrate Plasma (200 uL)
Sensitivity
10 pg/mL
Assay Range
31.3 - 1,000 pg/mL (Cell Culture Supernates, Serum, Platelet-poor EDTA Plasma, Platelet-poor Heparin Plasma, Platelet-poor Citrate Plasma)
Specificity
Natural and recombinant human GRO alpha.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 41.8 | 140 | 373 | 48.7 | 157 | 379 |
| Standard Deviation | 0.91 | 5.98 | 15.7 | 2.9 | 12.3 | 29.8 |
| CV% | 2.2 | 4.3 | 4.2 | 6 | 7.8 | 7.8 |
Serum, Platelet-poor EDTA Plasma, Platelet-poor Heparin Plasma, Platelet-poor Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 61.4 | 251 | 736 | 62.8 | 252 | 734 |
| Standard Deviation | 2.2 | 6 | 20 | 3.6 | 11.9 | 38.9 |
| CV% | 3.6 | 2.4 | 2.7 | 5.7 | 4.7 | 5.3 |
Recovery
The recovery of human GRO alpha spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 99 | 92-105 |
| Platelet-poor Citrate Plasma (n=4) | 90 | 80-98 |
| Platelet-poor EDTA Plasma (n=4) | 93 | 86-102 |
| Platelet-poor Heparin Plasma (n=4) | 95 | 83-109 |
| Serum (n=4) | 98 | 87-108 |
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有
Copyright © 深圳子科生物科技有限公司版权所有