返回 产品中心
Human M-CSF Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human M-CSF Quantikine ELISA Kit Summary
Catalog Size / Price
DMC00B 1 Kit / Please InquireSMC00B 1 Pack (6 Plates) / Please Inquire
PDMC00B 1 Pack (50 Plates) / Please Inquire
Assay Length
4.5 hours
Specificity
Natural and recombinant human M-CSF
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Saliva, Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 248 | 751 | 1503 | 280 | 827 | 1620 |
| Standard Deviation | 19.4 | 16.1 | 49 | 30.2 | 76 | 87.7 |
| CV% | 7.8 | 2.1 | 3.3 | 10.7 | 9.2 | 5.4 |
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 357 | 951 | 1767 | 343 | 1019 | 1956 |
| Standard Deviation | 17.4 | 24.9 | 49.4 | 43.4 | 84.1 | 88.8 |
| CV% | 4.9 | 2.6 | 2.8 | 12.6 | 8.3 | 4.5 |
Recovery
The recovery of M-CSF spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 99 | 90-107 |
| EDTA Plasma (n=4) | 100 | 93-106 |
| Heparin Plasma (n=4) | 99 | 93-106 |
| Serum (n=4) | 101 | 96-106 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of M-CSF were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有