返回 产品中心
Human Thrombospondin-2 Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human Thrombospondin-2 Quantikine ELISA Kit Summary
货号:DTSP20 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (13 uL), EDTA Plasma (13 uL), Heparin Plasma (13 uL), Human Milk (50 uL)
Sensitivity
0.068 ng/mL
Assay Range
0.3 - 20 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Human Milk)
Specificity
Natural and recombinant human Thrombospondin-2
Cross-reactivity
Cross-reactivity observed with 1 or more available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 4.08 | 7.73 | 10.9 | 4.11 | 7.71 | 11.4 |
| Standard Deviation | 0.189 | 0.229 | 0.281 | 0.295 | 0.511 | 0.726 |
| CV% | 4.6 | 3 | 2.6 | 7.2 | 6.6 | 6.4 |
Serum, EDTA Plasma, Heparin Plasma, Human Milk
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 3.97 | 7.96 | 11.4 | 4.12 | 7.68 | 11.4 |
| Standard Deviation | 0.198 | 0.205 | 0.284 | 0.259 | 0.403 | 0.527 |
| CV% | 5 | 2.6 | 2.5 | 6.3 | 5.2 | 4.6 |
Recovery
The recovery of Thrombospondin-2 spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 101 | 91-108 |
| EDTA Plasma (n=4) | 93 | 86-105 |
| Heparin Plasma (n=4) | 92 | 86-103 |
| Serum (n=4) | 93 | 86-105 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Thrombospondin-2 were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705