返回 产品中心
Human Pro-IL-1 beta/IL-1F2 Quantikine ELISA Kit
Human Pro-IL-1 beta/IL-1F2 Quantikine ELISA Kit Summary
货号:DLBP00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (200 uL), Serum (200 uL), EDTA Plasma (200 uL), Heparin Plasma (200 uL)
Sensitivity
8.9 pg/mL
Assay Range
23.4 - 1,500 pg/mL (Cell Culture Supernates), 31.2 - 2,000 pg/mL (Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human pro-IL-1 beta
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 85.5 | 240 | 432 | 85.1 | 248 | 457 |
| Standard Deviation | 4.7 | 11.5 | 20.8 | 6.2 | 14 | 28.3 |
| CV% | 5.5 | 4.8 | 4.8 | 7.3 | 5.6 | 6.2 |
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 126 | 374 | 670 | 121 | 342 | 626 |
| Standard Deviation | 6.4 | 15.7 | 34.5 | 8.8 | 17.9 | 34.9 |
| CV% | 5.1 | 4.2 | 5.1 | 7.3 | 5.2 | 5.6 |
Recovery
The recovery of pro-IL-1 beta spiked to three different levels throughout the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 108 | 99-115 |
| EDTA Plasma (n=5) | 103 | 95-108 |
| Heparin Plasma (n=5) | 101 | 92-109 |
| Serum (n=5) | 102 | 97-108 |
Linearity
To assess the linearity of the assay, five samples spiked with high concentrations of pro-IL-1 beta were diluted in the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有
Copyright © 深圳子科生物科技有限公司版权所有