Human Kallikrein 3/PSA Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human Kallikrein 3/PSA Quantikine ELISA Kit Summary
货号:DKK300 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL)
Sensitivity
0.069 ng/mL
Assay Range
0.9 - 60 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human KLK3/PSA
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 12.3 | 24.3 | 44.1 | 11.8 | 24.2 | 43.8 |
| Standard Deviation | 0.7 | 1.3 | 2 | 0.9 | 0.5 | 0.8 |
| CV% | 5.7 | 5.4 | 4.6 | 7.7 | 5.8 | 5.5 |
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 6.96 | 15.5 | 30 | 7.38 | 16.7 | 33 |
| Standard Deviation | 0.5 | 0.7 | 0.9 | 0.5 | 0.9 | 1.6 |
| CV% | 7.8 | 4.5 | 3 | 6.4 | 5.3 | 5 |
Recovery
The recovery of KLK3/PSA spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 99 | 93-105 |
| EDTA Plasma (n=4) | 100 | 93-103 |
| Heparin Plasma (n=4) | 101 | 97-109 |
| Serum (n=4) | 102 | 97-109 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of KLK3/PSA were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.