Human SP-D Quantikine ELISA Kit
Human SP-D Quantikine ELISA Kit Summary
货号:DSFPD0
规格:96T
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
5.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL)
Sensitivity
0.37 ng/mL
Assay Range
0.6 - 40 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human SP-D
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 6.07 | 11.7 | 23.4 | 5.63 | 11.2 | 22.8 |
| Standard Deviation | 0.495 | 0.588 | 1.46 | 0.495 | 0.999 | 1.78 |
| CV% | 8.2 | 5 | 6.2 | 8.8 | 8.9 | 7.8 |
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 5.99 | 11.3 | 24.3 | 6.63 | 13.1 | 26.9 |
| Standard Deviation | 0.371 | 0.921 | 1.63 | 0.617 | 1.14 | 2.5 |
| CV% | 6.2 | 8.2 | 6.7 | 9.3 | 8.7 | 9.3 |
Recovery
The recovery of SP-D spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 102 | 95-109 |
| EDTA Plasma (n=4) | 110 | 99-120 |
| Heparin Plasma (n=4) | 104 | 85-118 |
| Serum (n=4) | 106 | 89-118 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of SP-D were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.