返回 产品中心

Human Galectin-9 Quantikine ELISA Kit

    

 深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!

Human Galectin-9 Quantikine ELISA Kit Summary

货号:DGAL90 

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (25 uL), EDTA Plasma (25 uL), Heparin Plasma (25 uL), Saliva (100 uL), Urine (10 uL), Human Milk (10 uL)
Sensitivity
0.028 ng/mL
Assay Range
0.2 - 10 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Saliva, Urine, Human Milk)
Specificity
Natural and recombinant human Galectin-9.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

Product Summary

The Quantikine Human Galectin-9 Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human Galectin-9 in cell culture supernates, serum, plasma, saliva, urine, and human milk. It contains E. coli-expressed recombinant human Galectin-9 and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human Galectin-9 showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring human Galectin-9.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components

Cell Culture Supernates, Saliva, Human Milk

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean 1.34 3.58 6.13 1.44 3.43 6.05
Standard Deviation 0.05 0.12 0.22 0.08 0.18 0.32
CV% 3.7 3.4 3.6 5.6 5.2 5.3

Serum, EDTA Plasma, Heparin Plasma, Urine

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean 1.44 3.77 6.29 1.4 3.28 5.84
Standard Deviation 0.06 0.07 0.08 0.11 0.2 0.3
CV% 4.2 1.9 1.3 7.9 6.1 5.1

Recovery

The recovery of human Galectin-9 spiked to levels throughout the range of the assay was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 96 84-104
EDTA Plasma (n=4) 103 85-117
Heparin Plasma (n=4) 104 89-117
Serum (n=4) 106 86-117

Linearity

To assess the linearity of the assay, samples containing high concentrations of human Galectin-9 were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Human Galectin-9 ELISA Linearity

上一页

Human Fas/TNFRSF6/CD95 Quantikine ELISA Kit

下一页

Human t-Plasminogen Activator/tPA Quantikine ELISA Kit

电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有