Human GM-CSF Quantikine HS ELISA Kit
Human GM-CSF Quantikine HS ELISA Kit Summary
货号:HSGM0
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
6.5 hours
Sample Type & Volume Required Per Well
Serum (150 uL), EDTA Plasma (150 uL), Heparin Plasma (150 uL), Urine (200 uL)
Sensitivity
0.26 pg/mL
Assay Range
1.0 - 64 pg/mL (Serum, EDTA Plasma, Heparin Plasma), 0.5 - 32 pg/mL (Urine)
Specificity
Natural and recombinant human GM-CSF
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 2.69 | 7.05 | 24.8 | 3.77 | 14.3 | 27.4 |
| Standard Deviation | 0.14 | 0.33 | 0.73 | 0.37 | 1.17 | 1.85 |
| CV% | 5.2 | 4.7 | 2.9 | 9.8 | 8.2 | 6.8 |
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 18 | 18 | 18 |
| Mean | 3.68 | 22 | 44.6 | 2.8 | 16.2 | 33.2 |
| Standard Deviation | 0.35 | 1.07 | 1.76 | 0.66 | 2.47 | 4.6 |
| CV% | 9.5 | 4.9 | 4 | 23.5 | 15.3 | 13.9 |
Recovery
The recovery of GM-CSF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| EDTA Plasma (n=4) | 114 | 88-140 |
| Heparin Plasma (n=4) | 109 | 91-122 |
| Serum (n=4) | 109 | 94-128 |
| Urine (n=5) | 102 | 95-116 |
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of GM-CSF in various matrices and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.