Human LIGHT/TNFSF14 Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human LIGHT/TNFSF14 Quantikine ELISA Kit Summary
货号:DLIT00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL), Saliva (25 uL)
Sensitivity
16.5 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Saliva)
Specificity
Natural and recombinant human LIGHT
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 350 | 712 | 1388 | 387 | 754 | 1469 |
| Standard Deviation | 12 | 31.7 | 63.3 | 19.6 | 31.8 | 63.4 |
| CV% | 3.4 | 4.5 | 4.6 | 5.1 | 4.2 | 4.3 |
Cell Culture Supernates, Saliva
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 308 | 626 | 1181 | 318 | 616 | 1177 |
| Standard Deviation | 11.7 | 25.3 | 34.6 | 27 | 44.2 | 81.8 |
| CV% | 3.8 | 4 | 2.9 | 8.5 | 7.2 | 6.9 |
Recovery
The recovery of LIGHT spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 105 | 100-113 |
| EDTA Plasma (n=5) | 106 | 91-114 |
| Heparin Plasma (n=5) | 101 | 89-115 |
| Serum (n=5) | 102 | 85-114 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of LIGHT were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.