Human C-Peptide Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human C-Peptide Quantikine ELISA Kit Summary
货号:DICP00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
2.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL), Urine (10 uL)
Sensitivity
2.88 pmol/L
Assay Range
28.9 - 1,850 pmol/L (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine)
Specificity
Natural and synthetic human Insulin C-Peptide.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Intra-assay Precision (Precision within an assay)<br>Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Inter-assay Precision (Precision between assays)<br>Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 129 | 436 | 845 | 127 | 437 | 861 |
| Standard Deviation | 5.77 | 14.3 | 13.2 | 12.4 | 33 | 65.3 |
| CV% | 4.5 | 3.3 | 1.6 | 9.8 | 7.6 | 7.6 |
Recovery
The recovery of human Insulin C-Peptide spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 99 | 84-106 |
| EDTA Plasma (n=4) | 101 | 95-108 |
| Heparin Plasma (n=4) | 102 | 90-115 |
| Serum (n=4) | 103 | 96-109 |
| Urine (n=4) | 97 | 81-106 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human Insulin C-Peptide were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.