返回 产品中心
Human ACE Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human ACE Quantikine ELISA Kit Summary
货号:DACE00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), Heparin Plasma (10 uL), Saliva (25 uL)
Sensitivity
0.051 ng/mL
Assay Range
0.8 - 50 ng/mL (Cell Culture Supernates, Serum, Heparin Plasma, Saliva)
Specificity
Natural and recombinant human ACE. In addition to soluble ACE, this kit recognizes membrane-bound forms from lysates of cultured cells and supernates.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Saliva
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 3.6 | 10.7 | 22.1 | 3.44 | 10.3 | 21.8 |
| Standard Deviation | 0.08 | 0.3 | 0.95 | 0.21 | 0.4 | 0.87 |
| CV% | 2.2 | 2.8 | 4.3 | 6.1 | 3.9 | 4 |
Serum, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 3.57 | 10.9 | 22.5 | 3.77 | 11.7 | 24.4 |
| Standard Deviation | 0.12 | 0.43 | 0.9 | 0.29 | 0.6 | 1.2 |
| CV% | 3.4 | 3.9 | 4 | 7.7 | 5.1 | 4.9 |
Recovery
The recovery of ACE spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Supernates (n=4) | 98 | 89-106 |
| Heparin Plasma (n=4) | 100 | 94-106 |
| Serum (n=4) | 99 | 91-106 |
Linearity
To assess the linearity of the assay, samples containing high concentrations of ACE were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有
Copyright © 深圳子科生物科技有限公司版权所有