Human Uteroglobin Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human Uteroglobin Quantikine ELISA Kit Summary
货号:DUGB00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL), Saliva (50 uL), Urine (50 uL), Human Milk (50 uL)
Sensitivity
0.217 ng/mL
Assay Range
0.8 - 50 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Saliva, Urine, Human Milk)
Specificity
Natural and recombinant human Uteroglobin
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Saliva, Urine, Human Milk
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 5.1 | 15.4 | 31.3 | 5.2 | 15.4 | 29.9 |
| Standard Deviation | 0.29 | 0.34 | 1.04 | 0.31 | 0.72 | 2.25 |
| CV% | 5.7 | 2.2 | 3.3 | 6 | 4.7 | 7.5 |
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 5.2 | 15.8 | 31.7 | 5 | 14.7 | 28.8 |
| Standard Deviation | 0.21 | 0.49 | 1.16 | 0.28 | 0.92 | 2.09 |
| CV% | 4 | 3.1 | 3.7 | 5.6 | 6.3 | 7.3 |
Recovery
The recovery of Uteroglobin spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 102 | 99-109 |
| EDTA Plasma (n=4) | 95 | 85-107 |
| Heparin Plasma (n=4) | 95 | 86-109 |
| Serum (n=4) | 95 | 85-104 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Uteroglobin were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.