返回 产品中心
Human PDGF-DD Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human PDGF-DD Quantikine ELISA Kit Summary
货号:DDD00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), Platelet-poor EDTA Plasma (50 uL), Platelet-poor Heparin Plasma (50 uL), Saliva (50 uL), Human Milk (50 uL)
Sensitivity
3.51 pg/mL
Assay Range
31.3 - 2,000 pg/mL (Cell Culture Supernates, Serum, Platelet-poor EDTA Plasma, Platelet-poor Heparin Plasma, Saliva, Human Milk)
Specificity
Natural and recombinant human PDGF-DD
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, Platelet-poor EDTA Plasma, Platelet-poor Heparin Plasma, Saliva, Human Milk
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 176 | 510 | 986 | 198 | 559 | 1095 |
| Standard Deviation | 3.69 | 11.9 | 17.2 | 16 | 38.5 | 55.3 |
| CV% | 2.1 | 2.3 | 1.7 | 8.1 | 6.9 | 5.1 |
Recovery
The recovery of PDGF-DD spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 102 | 83-112 |
| Platelet-poor EDTA Plasma (n=4) | 93 | 83-109 |
| Platelet-poor Heparin Plasma (n=4) | 94 | 84-113 |
| Serum (n=4) | 101 | 86-112 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of PDGF-DD were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有
Copyright © 深圳子科生物科技有限公司版权所有