返回 产品中心
Human EGFR Quantikine ELISA Kit
Human EGFR Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DEGFR0 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL), Human Milk (50 uL)
Sensitivity
0.036 ng/mL
Assay Range
0.3 - 20 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Human Milk)
Specificity
Natural and recombinant human EGF R
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Human Milk
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 1.65 | 5.08 | 10.6 | 1.63 | 5 | 10 |
| Standard Deviation | 0.094 | 0.17 | 0.507 | 0.144 | 0.416 | 0.816 |
| CV% | 5.7 | 3.3 | 4.8 | 8.8 | 8.3 | 8.2 |
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 1.69 | 5.33 | 11.2 | 1.7 | 5.21 | 10.4 |
| Standard Deviation | 0.072 | 0.199 | 0.616 | 0.161 | 0.522 | 1.03 |
| CV% | 4.3 | 3.7 | 5.5 | 9.4 | 10 | 10 |
Recovery
The recovery of EGF R spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 98 | 86-106 |
| EDTA Plasma (n=4) | 99 | 91-108 |
| Heparin Plasma (n=4) | 101 | 92-111 |
| Serum (n=4) | 101 | 89-117 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of EGF R were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.