返回 产品中心
Human IL-2 Quantikine ELISA Kit
Human IL-2 Quantikine ELISA Kit Summary
货号:D2050
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL)
Sensitivity
7 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human IL-2
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 67.5 | 438 | 851 | 232 | 620 | 1148 |
| Standard Deviation | 2.9 | 14.4 | 26.9 | 23.2 | 34.4 | 45.5 |
| CV% | 4.3 | 3.3 | 3.2 | 10 | 5.5 | 4 |
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 98 | 619 | 1183 | 250 | 755 | 1470 |
| Standard Deviation | 4.2 | 18 | 24 | 12.6 | 27.8 | 74.1 |
| CV% | 4.3 | 2.9 | 2 | 5 | 3.7 | 5 |
Recovery
The recovery of human IL-2 spiked to three levels throughout the range of the assay was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 101 | 96-107 |
| Citrate Plasma (n=5) | 105 | 98-111 |
| EDTA Plasma (n=5) | 118 | 104-139 |
| Heparin Plasma (n=5) | 109 | 94-129 |
| Serum (n=5) | 99 | 89-114 |
Linearity
To assess the linearity of the assay, five samples were spiked with high concentrations of IL-2 in various matrices and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705
Copyright © 深圳子科生物科技有限公司版权所有