返回 产品中心
Human CCL5/RANTES Quantikine ELISA Kit
Human CCL5/RANTES Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DRN00B
Assay Length
3.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL), Urine (100 uL)
Sensitivity
6.6 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma, Urine)
Specificity
Natural and recombinant human RANTES
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 91.9 | 573 | 1120 | 92 | 511 | 1105 |
| Standard Deviation | 2.3 | 10 | 32.4 | 6.2 | 32.7 | 70.7 |
| CV% | 2.5 | 1.7 | 2.9 | 6.7 | 6.4 | 6.4 |
EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 108 | 599 | 1191 | 88 | 494 | 1063 |
| Standard Deviation | 3.9 | 14.3 | 41.5 | 9.1 | 43.1 | 89.6 |
| CV% | 3.6 | 2.4 | 3.5 | 10.3 | 8.7 | 8.4 |
Recovery
The recovery of RANTES spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 101 | 95-110 |
| Platelet-poor Citrate Plasma (n=4) | 99 | 93-105 |
| Platelet-poor EDTA Plasma (n=4) | 98 | 93-103 |
| Platelet-poor Heparin Plasma (n=4) | 99 | 93-105 |
| Serum (n=5) | 101 | 94-107 |
| Urine (n=10) | 102 | 88-114 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of RANTES were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.