Human PlGF Quantikine ELISA Kit
Human PlGF Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DPG00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours or 4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL), Urine (100 uL)
Sensitivity
7 pg/mL
Assay Range
15.6 - 1,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma, Urine)
Specificity
Natural and recombinant human PlGF
Cross-reactivity
Cross-reactivity observed with 1 or more available related molecules.Cross-species reactivity not tested.
Interference
Interference observed with 1 or more available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 40.9 | 138 | 541 | 46.9 | 149 | 555 |
| Standard Deviation | 2.4 | 7.3 | 38.2 | 6.4 | 13.5 | 32.9 |
| CV% | 5.9 | 5.3 | 7.1 | 13.6 | 9.1 | 5.9 |
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma, Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 54.3 | 175 | 658 | 55 | 184 | 724 |
| Standard Deviation | 3.8 | 6.3 | 36.9 | 6.5 | 20.3 | 78.9 |
| CV% | 7 | 3.6 | 5.6 | 11.8 | 11 | 10.9 |
Recovery
The recovery of PlGF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 97 | 82-107 |
| Citrate Plasma (n=5) | 99 | 93-120 |
| EDTA Plasma (n=5) | 102 | 92-117 |
| Heparin Plasma (n=5) | 94 | 81-113 |
| Serum (n=5) | 96 | 88-118 |
| Urine (n=4) | 97 | 85-112 |
Linearity
To assess the linearity of the assay, five samples were spiked with high concentrations of PlGF in various matrices and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.