Human PDGF-BB Quantikine ELISA Kit Summary
Human PDGF-BB Quantikine ELISA Kit Summary
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DBB00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (10 uL), Platelet-poor EDTA Plasma (100 uL), Platelet-poor Heparin Plasma (100 uL)
Sensitivity
15 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human PDGF-BB
Cross-reactivity
Cross-reactivity observed with 1 or more available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 310 | 759 | 1093 | 287 | 772 | 1082 |
| Standard Deviation | 14 | 18.6 | 25.5 | 21.9 | 56.9 | 89.8 |
| CV% | 4.5 | 2.4 | 2.3 | 7.6 | 7.4 | 8.3 |
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 268 | 639 | 935 | 251 | 655 | 918 |
| Standard Deviation | 11.3 | 28.4 | 26.8 | 22.4 | 38.5 | 59.3 |
| CV% | 4.2 | 4.4 | 2.9 | 8.9 | 5.9 | 6.5 |
Recovery
The recovery of PDGF-BB spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 94 | 86-102 |
| Platelet-poor EDTA Plasma (n=6) | 101 | 88-115 |
| Platelet-poor Heparin Plasma (n=6) | 103 | 89-112 |
| Serum (n=6) | 102 | 93-114 |
Linearity
To assess the linearity of the assay, samples containing or spiked with high concentrations of PDGF-BB were diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.