返回 产品中心
Human SCF Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human SCF Quantikine ELISA Kit Summary
Catalog #DCK00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL)
Sensitivity
9 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human SCF. This assay cross-reacts 15.6% with Canine SCF, 16.1% with Feline SCF, 3.5% with Mouse SCF and 1.0% with Rat SCF.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity observed with 1 or more species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 139 | 655 | 1419 | 142 | 703 | 1574 |
| Standard Deviation | 5.5 | 12.9 | 54 | 14.6 | 36 | 94 |
| CV% | 4 | 2 | 3.8 | 10.3 | 5.1 | 6 |
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 153 | 709 | 1427 | 151 | 729 | 1552 |
| Standard Deviation | 3.6 | 11.6 | 61 | 12.1 | 45.8 | 125 |
| CV% | 2.4 | 1.6 | 4.3 | 8 | 6.3 | 8.1 |
Recovery
The recovery of SCF spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=2) | 85 | 75-93 |
| Citrate Plasma (n=8) | 94 | 84-112 |
| EDTA Plasma (n=8) | 98 | 82-110 |
| Heparin Plasma (n=8) | 99 | 88-116 |
| Serum (n=8) | 99 | 87-111 |
Linearity
To assess the linearity of the assay, the following biological samples were diluted with the appropriate Calibrator Diluent and then assayed. The cell culture media was spiked with exogenous SCF.