Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL), Saliva (10 uL), Urine (25 uL), Human Milk (10 uL)
Sensitivity
0.54 ng/mL
Assay Range
1.6 - 100 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Saliva, Urine, Human Milk)
Specificity
Natural and recombinant human Progranulin. Trypsin and Elastase cleaved Progranulin show reduced recognition in this assay, demonstrating specificity for Progranulin.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Saliva
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 13.5 | 32.7 | 64.8 | 14.4 | 33.1 | 64.4 |
| Standard Deviation | 0.7 | 1.9 | 4.1 | 1 | 2.3 | 4.4 |
| CV% | 5.2 | 5.8 | 6.3 | 6.9 | 6.9 | 6.8 |
Serum, EDTA Plasma, Heparin Plasma, Urine, Human Milk
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean | 15.5 | 36.6 | 70.8 | 15.7 | 35.8 | 69.1 |
| Standard Deviation | 0.9 | 1.7 | 3.1 | 1.3 | 2.9 | 5.1 |
| CV% | 5.8 | 4.6 | 4.4 | 8.3 | 8.1 | 7.4 |
Recovery
The recovery of Progranulin spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 107 | 103-112 |
| EDTA Plasma (n=4) | 102 | 95-110 |
| Heparin Plasma (n=4) | 100 | 92-108 |
| Serum (n=4) | 104 | 90-115 |
| Urine (n=4) | 109 | 99-116 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Progranulin were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.