返回 产品中心
Porcine IFN-gamma Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Porcine IFN-gamma Quantikine ELISA Kit Summary
货号:PIF00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL)
Sensitivity
11.2 pg/mL
Assay Range
39.0 - 2,500 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant porcine IFN-gamma
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 29 | 34 | 34 |
| Mean | 124 | 272 | 913 | 120 | 282 | 904 |
| Standard Deviation | 5.8 | 7.4 | 26.5 | 14 | 15.7 | 84.7 |
| CV% | 4.7 | 2.7 | 2.9 | 11.7 | 5.6 | 9.4 |
Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 36 | 32 | 36 |
| Mean | 226 | 449 | 1518 | 227 | 515 | 1522 |
| Standard Deviation | 10.1 | 12 | 40.4 | 17.1 | 51.3 | 97.7 |
| CV% | 4.5 | 2.7 | 2.7 | 7.5 | 10 | 6.4 |
Recovery
The recovery of porcine IFN-gamma spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Supernates (n=4) | 97 | 84-116 |
| EDTA Plasma (n=5) | 98 | 85-111 |
| Heparin Plasma (n=5) | 106 | 86-118 |
| Serum (n=5) | 98 | 81-119 |
Linearity
To assess the linearity of the assay, samples containing or spiked with high concentrations of porcine IFN-gamma in each matrix were diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.