Human ADAMTS13 Quantikine ELISA Kit Summary
货号:DADT130 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL), Citrate Plasma (10 uL)
Sensitivity
0.26 ng/mL
Assay Range
0.8 - 50 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human ADAMTS13. This kit also recognizes a truncated version of recombinant human ADAMTS13 (amino acids Gln34-Trp688; Accession # NP_620594).
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 4.78 | 13.8 | 29.7 | 4.51 | 14.2 | 30.2 |
| Standard Deviation | 0.177 | 0.364 | 0.685 | 0.256 | 0.722 | 1.08 |
| CV% | 3.7 | 2.6 | 2.3 | 5.7 | 5.1 | 3.6 |
Recovery
The recovery of ADAMTS13 spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=8) | 100 | 91-111 |
| Citrate Plasma (n=4) | 97 | 87-114 |
| EDTA Plasma (n=4) | 97 | 89-109 |
| Heparin Plasma (n=4) | 97 | 85-115 |
| Serum (n=4) | 97 | 85-106 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human ADAMTS13 were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.