返回 产品中心
Human Glypican 3 Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human Glypican 3 Quantikine ELISA Kit
| Catalog # | Availability | Size / Price | |
|---|---|---|---|
| DGLY30 | Contact your distributor for availability. | 1 Kit / Please Inquire | |
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Cell Lysates (20 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL)
Sensitivity
20.6 pg/mL
Assay Range
78.1 - 5,000 pg/mL (Cell Culture Supernates, Cell Lysates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human Glypican 3.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.<br>
Cell Culture Supernates, Cell Lysates, Serum, EDTA Plasma, Heparin Plasma
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 412 | 1208 | 2387 | 466 | 1231 | 2359 |
| Standard Deviation | 12 | 48 | 79.6 | 38.4 | 75.3 | 143 |
| CV% | 2.9 | 4 | 3.3 | 8.2 | 6.1 | 6.1 |
Recovery
The recovery of human Glypican 3 spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=4) | 99 | 88-112 |
| EDTA Plasma (n=4) | 88 | 81-95 |
| Heparin Plasma (n=4) | 89 | 81-96 |
| Serum (n=4) | 89 | 82-97 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human Glypican 3 were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
电话咨询:0755-33164177/13924667705