返回 产品中心
Human ADAM12 Quantikine ELISA Kit
深圳子科生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human ADAM12 Quantikine ELISA Kit
| Catalog # | Availability | Size / Price | |
|---|---|---|---|
| DAD120 | Contact your distributor for availability. | 1 Kit / Please Inquire | |
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (25 uL), Heparin Plasma (25 uL), Urine (50 uL)
Sensitivity
0.076 ng/mL
Assay Range
0.2 - 10 ng/mL (Cell Culture Supernates, Serum, Heparin Plasma, Urine)
Specificity
Natural and recombinant human ADAM12. This kit recognizes pro-ADAM12, mature ADAM12, and ADAM12/TIMP-3 complex.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, Heparin Plasma, Urine
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean | 1.39 | 3.57 | 6.84 | 1.43 | 3.62 | 6.74 |
| Standard Deviation | 0.057 | 0.131 | 0.235 | 0.079 | 0.145 | 0.299 |
| CV% | 4.1 | 3.7 | 3.4 | 5.5 | 4 | 4.4 |
Recovery
The recovery of human ADAM12 spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=8) | 98 | 88-105 |
| Heparin Plasma (n=4) | 97 | 92-106 |
| Serum (n=4) | 98 | 89-106 |
| Urine (n=4) | 99 | 91-106 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human ADAM12 were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.